There are two media functions and two alternative applications:
Media features
- Standard mode, intended for normal preparations.
- Chocolate agar, two-stage program with two heating ramps for the preparation of complex media. After the first sterilization phase, supplements, e.g. blood, can be added through the supplement port. Then a further heating phase takes place.
Applications
- Heat bath, to pre-treat the medium before sterilization with a selectable temperature between 30-80°C.
- Autoclave, with special insert valve it is possible to autoclave/thermostat liquids in containers.
Safe media handling
Before starting, a low pressure test is performed to check that the system is tight. In addition to electronic control of pressure and temperature, two additional mechanical safety levels are now available. The Mediaclave has a powerful magnet that changes direction during the preparation process. This means that a more homogeneous mixture is obtained.
Contact the product specialist:
Urban George
Urban.george@bergmanlabora.se
Tel: 08-625 18 12 |
Technical specification
Steel-free vessel 10 or 30 LL
Temperature range
- Sterilization 70°C – 122°C 70°C – 122°C
- Dispensing 25°C – 80°C 25°C – 80°C
- Water bath 30°C – 80°C 30°C – 80°C
Rotation speed: Selectable 50 – 200 rpm, reversing direction 50 – 200 rpm, reversing direction
Utilities
Interfaces 2 x RS232, Ethernet, USB port, AUX 2 x RS232, Ethernet, USB port, AUX contact, footswitch, ext. pinch valve contact, footswitch, ext. pinch valve